The InfraCal 2 TRANS-SP  is recommended for measuring oil in water, TPH in soil or FOG in wastewater concentration levels using the traditional EPA methods 413.2 and 418.1 or ASTM method D7066-04 and Freon-113, hydrocarbon-free grade of perchloroethylene, AK-225, S-316 or other spectroscopic infrared transparent solvent as the extracting solvent. Since the extract is deposited into a 10 mm quartz cuvette cell with Teflon stopper, light end volatile components are retained for measurement. Accurate down to 0.1 ppm.

The InfraCal 2 TRANS-SP, the latest introduction to our easy to use analyzers, uses the same measurement technology as the original InfraCal Model CVH, with added features such as data storage and transfer, multiple calibrations and password protection for instrument settings.

Specifications
       • MDL Water:  0.1ppm
       • MDL Soil:  1ppm
       • Solvent(s):  Perchloroethylene, S-316, Freon-113
       • Method Compares:  EPA 1664, ISO 9377-2
       • Method Complies:  ASTM D7066, EPA 413.2, EPA 418.1

Highlights

The fact that InfraCal Analyzers over 10 years old are still operating in the corrosive environment of an offshore oil rigs attests to their rugged, dependability.   The analysis procedure takes less than 15 minutes and can be done by minimally trained personnel.  They use the field proven infrared technology for measuring oil in water, oil content in drilling mud or drill cuttings, TPH (total petroleum hydrocarbons) in soil and FOG (fats, oil and grease) in wastewater.

Operating Principle

The InfraCal TOG/TPH Analyzer makes use of the fact that hydrocarbons such as oil and grease can be extracted from water or soil through the use of an appropriate solvent and extraction procedure.  The extracted hydrocarbons absorb infrared energy at a specific wavelength and the amount of energy absorbed is proportional to the concentration of oil and grease in the solvent.  The analyzer can be calibrated to read out directly in the desired units.

Calibrations

The InfraCal 2 Model TRANS-SP is available with a single or multiple calibrations. Analysis results of total oil and grease & total petroleum hydrocarbon concentration levels are displayed in parts per million (ppm) or milligram/kg.

PN  Factory Calibrations
405-2035-41 Includes a factory calibration using S-316  standards for samples containing primarily petroleum hydrocarbons with minimal animal/plant oils
TOG/TPH High Range S-316
Water: 10, 25, 50, 100, 150, and 200 mg/L
Soil: 100, 250, 500, 1000, 1500, and 2000 mg/kg
405-2035-8041  Includes a factory calibration using S-316  standards for samples containing primarily petroleum hydrocarbons with minimal animal/plant oils
TOG/TPH Low and High S-316
Water: 0.5, 1, 2.5, 5.0, and 7.5 mg/L
Water: 10, 25, 50, 100, 150, and 200 mg/L
Soil: 100, 250, 500, 1000, 1500, and 2000 mg/kg
405-2035-40  Includes a factory calibration using S-316  standards for samples containing primarily animal and/or plant oils
FOG/TOG High Range S-316

Water: 10, 25, 50, 100, 150, and 200mg/L
405-2035-78 Includes a factory calibration using Tetracloroethylene standards for samples containing primarily petroleum hydrocarbons with minimal animal/plant oils
TOG/TPH High Range Tetracloroethylene
Water: 10, 25, 50, 100, 150, and 200 mg/L
Soil: 100, 250, 500, 1000, 1500, and 2000 mg/kg
405-2035-7978  Includes a factory calibration using Tetracloroethylene standards for samples containing primarily petroleum hydrocarbons with minimal animal/plant oils
TOG/TPH Low and High Tetracloroethylene
Water: 0.5, 1, 2.5, 5.0, and 7.5 mg/L
Water: 10, 25, 50, 100, 150, and 200 mg/L
Soil: 100, 250, 500, 1000, 1500, and 2000 mg/kg

Calibration Standard Sets

PN Calibration Standards Sets
403-1080 TOG/TPH Low Range S-316 TOG/TPH Five Point Calibration Standard Set Light Mineral Oil in S-316 prepared by a certified laboratory. The set contains 6 standards from 0.5mg/L to 7.5 mg/L (ppm) for water and 5 mg/kg to 75 mg/kg for soil, includes solvent blanks. Recommended for customers in the petroleum industry who are primarily measuring low levels of petroleum hydrocarbons in water or soil.
403-1041 TOG/TPH High Range S-316 Six Point Calibration Standard Set: Light Mineral Oil in S-316 prepared by a certified laboratory for TOG/TPH Analysis. Nominal concentration values are 100 mg/kg (ppm), 250 mg/kg, 500 mg/kg, 1000 mg/kg, 1500 mg/kg and 2000 mg/kg for soil assuming a 1:1 extraction ratio; and 10 mg/L (ppm), 25 mg/L, 50 mg/L, 100 mg/L, 150 mg/L and 200 mg/L for water assuming a 10:1 extraction ratio. The set contains 6 standards along with a solvent blank in 4 ml flame sealed ampoules. Recommended for customers in the petroleum industry who are primarily measuring petroleum hydrocarbons in water or soil.
403-1040 FOG/TOG High Range S-316 TOG/FOG Six Point Calibration Standard Set: 50:50 Isooctane/Octanoic Acid in S-316 as per ASTM Method D 7066-04 prepared by a certified laboratory. Nominal concentration values are 10 mg/L (ppm), 25 mg/L, 50 mg/L, 100 mg/L, 150 mg/L and 200 mg/L for water assuming a 10:1 extraction ratio. The set contains 6 standards along with a solvent blank in 4 ml flame sealed ampoules. Recommended for customers measuring non-polar (petroleum) and polar (animal and vegetable fats) hydrocarbons for TOG or FOG measurements such as municipal wastewater or industrial wastewater that includes animal and vegetable fats.
403-1079 TOG/TPH Low Range Tetrachloroethylene TOG/TPH Five Point Calibration Standard Set Light Mineral Oil in Tetrachloroethylene prepared by a certified laboratory. The set contains 6 standards from 0.5mg/L to 7.5 mg/L (ppm) for water and 5 mg/kg to 75 mg/kg for soil, includes solvent blanks. Recommended for customers in the petroleum industry who are primarily measuring low levels of petroleum hydrocarbons in water or soil.
403-1078 TOG/TPH High Range Tetrachloroethylene TOG/TPH Six Point Calibration Standard Set: Light Mineral Oil in Tetrachloroethylene prepared by a certified laboratory for TOG/TPH Analysis. Nominal concentration values are 100 mg/kg (ppm), 250 mg/kg, 500 mg/kg, 1000 mg/kg, 1500 mg/kg and 2000 mg/kg for soil assuming a 1:1 extraction ratio; and 10 mg/L (ppm), 25 mg/L, 50 mg/L, 100 mg/L, 150 mg/L and 200 mg/L for water assuming a 10:1 extraction ratio. The set contains 6 standards along with a solvent blank in 4 ml flame sealed ampoules. Recommended for customers in the petroleum industry who are primarily measuring petroleum hydrocarbons in water or soil.

Accessories

PN Accessories
403-0013 Carrying Case for InfraCal Analyzers with pre-diced pluck foam.
18.5” x  14.6” x 8”; 10 lbs.
403-1086 Dust Cover
403-0012 Set of Four 10mm Quartz Cuvette Cells with Teflon Stopper
 
Field Sampling Kits
 
PN Field Sampling Kits
403-1063 Field Sampling Kit for TOG-TPH in Water
100ml Graduated Cylinder
25ml Graduated Cylinder
125 ml Wash Bottle
20ml Beaker
100ul Syringe
Disposable Pipettes (100 pc.) 
Glass Funnel
177ml Graduated Bottles (48 pc)
Septa Caps (48 pc)
Sodium Sulfate, 500g
Silica Gel, 500g
Filter Paper, 100pc
20ml Luer-Lok Syringe & Tip (100pc)
403-1010 Field Sampling Kit for TOG-TPH in Soil
100ml Graduated Cylinder
25ml Graduated Cylinder
125 ml Wash Bottle
20ml Beaker
100ul Syringe
Disposable Pipettes (100 pc.)
Glass Funnel
40ml Vials (72 pc.)
Pocket Balance
Sample Spatula
Silica Gel, 500g
Filter Paper, 100pc
60ml Syringe Cartridge & Plunger (100 pc)
403-1083 Field Sampling Kit for Sub-ppm in Water
10ml Glass Luer-Lok Syringe
Filters for Glass Luer-Lok Syringe (100 pc)
This kit is to be used in addition to 403-1063 for sub-ppm in water sampling
 
Specifications
 
Item Specifications
Analytical Wavelength/Wave Number 3.4 µm, 2930 cm-1
Power Requirements 18 volts DC, 3.3 amps, internal battery pack available
Power Supply Universal AC/DC provided
Weight 5.8 lb. (2.6 kg) with battery -7.0 lbs (3.2 kg)
Dimensions (17 cm) x 7.8" (19.8 cm) x 5.2" (13.2 cm)
Suggested Operating Range 40°F (5°C) – 110°F (40°C)
Measurement Range  
For Water 0.1-1000+ ppm
For Soil 1-5000+ ppm
Analysis Time 10-15 minutes, including extraction process
Instrument Repeatability ±0.1 ppm
Communication Port USB, RS 232